Suspended Solids Analyzer (Online)

Suspended Solids Analyzer - TCS1000 / M100

Built in auto cleaning wiper with hard-to-scratch sapphire glass makes maintenance easy with reliable & stable measurement in harsh conditions.

Suspended solids transmitter uses a method of 90°C scattered light. In the method, a light source of a Detector illuminates the surfaces of suspended matter floating in the water and the light is scattered by these surfaces is detected by a photo element installed at an angle of 90° from the light source. The suspended solid is determined by the amount of scattered light.

Suspended Solid Analyzer Working Principle


Features

• Build-in wiping cleaning system certainly keeps lenses clean

• Check and clean the window surface once every month.

• “Easy to Clean” scratch-proof sapphire glass for the optical windows

• Compact size can fit into any installation site

• Monitoring of clarifier overflow weirs in waste water treatment plants

• Final outlet of effluent from dissolved air floatation plants

• Raw water inlet measurements in water treatment plants

• Surface water monitoring

• Solids loading in rivers and streams

• Product breakthrough on plate heat exchangers


Applications

• Compressed Air

• Breathing Air

• Beverage Industry

• Automotive

• General Engineering

• Drying of Plastics

• Military / Defense

• Research & Development

• General Gases


Specifications

Technical Details
Sensor TCS-1000
Measuring Range 0-1000 (mg/l) (standard substance:kaolin)
Output Analog (4-20mA) signal output:
resistance load of 300Ω (Max.)
4mA: 0 (mg/l)
20mA: 1000 (mg/l)
Self-checking output:
open collector (12VDC 20mA Max.)
Input Calibration signal input
Cleaning System Both sided swing wiper cleaning system
Operating Temperature 0-40°C (unfrozen)
Material SUS 316L, sapphire glass, fluorocarbon rubber, EPDM, POM
Degree of Protection IP 68, maximum depth of 2 meters (underwater type)
Detector Cable Length 10m
Transmitter M100
Power 110 - 240 VAC 50/60Hz
Display Resolution 0.1mg/l(0-49.9mg/1)
1mg/l(50-1000mg/l)
Output Calibration Output
Power Supply for Sensor (12 VDC)
Signal Output
(analog 4-20mA, resistance load of 300Ω or less)
Self-checking Relay Output (240 VAC, 1A)
Alarm Relay Output (240 VAC, 1A)
Input (Detector) Detector signal input (analog 4-20mA, input resistance approx.100)
Self-checking input
Operating Temperature -20° to +50°C
Material Polycarbonate
Protection IP65
Note: Specifications are subject to change witout notice.

Enquiry

For all inquiries, you may contact us through Telephone, Email or Live Chat Support.

Operating Hours

Monday to Friday: 8.00am to 5.30pm

Technical Support (After Office Hours)

Tel: (65) 9876 6258

Contact Information

By Phone / FaxBy Email
Tel: (65) 6553 0366
Fax: (65) 6553 0828
info@flotech.com.sg

Flotech Approvals